Standard Operating Procedure: Syphilis Serology (VDRL / RPR)
1. Purpose and Scope
To perform nontreponemal serologic screening for syphilis (VDRL and RPR) and to quantitate reactive titers for monitoring disease activity and treatment response, with a protocol for confirmation by treponemal tests.
2. Specimen Requirements
- Type: Serum (plain tube or SST); fresh, non-hemolyzed.
- Volume: Minimum 1.0 mL.
- Interference: Lipemia and hemolysis interfere with flocculation reading; grossly contaminated sera cause false reactions.
- Stability: Serum stable 3-5 days at 2-8°C; freeze if testing is delayed beyond the validated window.
- Timing: RPR uses warm serum; do not heat inactivate because heat destroys some antibodies and destabilizes the reagent.
3. Safety and Precautions
- Standard PPE (gloves, lab coat). Serum and syphilis reagents are potentially infectious.
- Dispose of pipettes and slides as biohazard; do not mouth-pipette.
- RPR/VDRL reagents contain cardiolipin and are alcohol-based; keep away from open flames.
4. Equipment and Reagents
- VDRL flocculation slide or RPR card kit with RPR antigen (carbonized cardiolipin-lectinin-cholesterol).
- Rotation platform (rotator) at the specified speed and serum dilution pipettes.
- Syringe or dropper delivering the correct antigen drop size.
- Positive (weakly reactive) and negative controls provided with the kit.
5. Step-by-Step Procedure
- Bring serum and RPR card/antigen to room temperature; run weakly reactive positive and negative controls before patient samples.
- Place the required amount of serum (or dilution) on the card ring.
- Add one drop of antigen, mix with the applicator, and rotate at the specified speed for the validated time.
- Read under the card lamp for clumping: record as nonreactive, weak reactive, or reactive.
- Titrate reactive sera by twofold dilution (1:2, 1:4, 1:8 ... 1:64 and higher) and report the endpoint titer.
- For VDRL, use the flocculation slide method with fresh serum; titrate reactive samples similarly.
- Consult the algorithm for treponemal confirmation (TP-PA/FTA-ABS/EIA) on any reactive result before reporting a positive diagnosis.
6. Quality Control
- The weakly reactive control must give a weak positive pattern in the expected low-titer range; the negative control must remain nonreactive.
- Check antigen uniformity (a flocculent, slightly clumped preparation is discarded).
- Participate in EQA for syphilis serology; validate batch of cards and rotator timing.
7. Decision limits
The interpretive thresholds below are clinical decision limits (there is no canonical reference range for serologic reagents).
- Reactive screen: any reactive (or weak reactive) nontreponemal result must be confirmed with a treponemal test (TP-PA, FTA-ABS, or EIA/CMIA) — a reactive screen alone is never reported as syphilis.
- Biological false positives (BFP): nnon-reactive treponemal test with a reactive nontreponemal screen occurs in pregnancy, autoimmune disease, intravenous drug use, infectious mononucleosis, malaria, and advanced age.
- Titer monitoring: a change of fourfold (two dilutions) between serial titers is considered clinically significant; a falling titer indicates adequate treatment response, a rising titer suggests progression or reinfection.
- Prozone effect: very high antibody titers (secondary syphilis) can produce a false-negative screen; dilute the serum and repeat when clinical suspicion is high.
- Primary syphilis: the nontreponemal screen may be nonreactive in early primary disease because antibody has not yet appeared.