Standard Operating Procedure: D-Dimer Testing
1. Purpose and Scope
This procedure defines the step-by-step method for performing D-dimer testing to aid in the diagnosis and exclusion of venous thromboembolism (VTE), disseminated intravascular coagulation (DIC), and other thrombotic disorders.
2. Specimen Requirements
- Specimen Type: Platelet-poor plasma.
- Anticoagulant: 3.2% (0.109 M) sodium citrate (blue top tube).
- Collection: Fill tube to proper volume (9:1 blood-to-anticoagulant ratio).
- Volume: Minimum 2 mL whole blood (yields approximately 1 mL plasma).
- Processing: Centrifuge within 2 hours of collection at 1500-2000 × g for 15 minutes at room temperature.
- Storage/Stability:
- Room temperature: Test within 4 hours
- Refrigerated (2-8°C): Stable for 24 hours
- Frozen (-20°C or below): Stable for 1 month (avoid repeated freeze-thaw cycles)
- Rejection Criteria: Underfilled tubes, clotted samples, hemolyzed or lipemic samples, improper anticoagulant.
3. Safety and Precautions
- Wear appropriate PPE (lab coat, gloves, eye protection).
- Treat all specimens as potentially infectious.
- Handle reagents according to manufacturer's safety data sheets.
- Dispose of waste in appropriate biohazard containers.
- Never mouth pipette.
4. Equipment and Reagents
- Equipment: Automated coagulation analyzer or ELISA reader, centrifuge, micropipettes, timer, vortex mixer, 37°C water bath or heating block (if required).
- Reagents: D-dimer test kit (latex-enhanced immunoturbidimetric, ELISA, or fluorescence immunoassay), calibration standards, quality control materials, citrate diluent buffer.
5. Step-by-Step Procedure
A. Specimen Collection and Processing
- Perform venipuncture using standard technique.
- Collect blood into 3.2% sodium citrate tube, filling to the indicated volume.
- Gently invert the tube 3-4 times to mix.
- Transport to laboratory at room temperature.
- Centrifuge at 1500-2000 × g for 15 minutes within 2 hours of collection.
- Carefully remove plasma without disturbing the cell layer.
- If not testing immediately, aliquot and store appropriately.
B. Instrument Setup and Calibration
- Turn on the analyzer and allow to warm up per manufacturer's instructions.
- Load D-dimer reagents, calibrators, and controls.
- Perform calibration according to manufacturer's schedule (typically with each new reagent lot or every 30 days).
- Verify calibration is acceptable before proceeding.
C. Quality Control
- Run at least two levels of quality control material with each batch of patient samples.
- Controls should be run:
- At the beginning of each shift
- With each new reagent lot
- After instrument maintenance
- As per laboratory QC policy
- Verify control results are within acceptable ranges before reporting patient results.
- Document all QC results.
D. Testing Procedure (Immunoturbidimetric Method)
- Bring all reagents and samples to room temperature (if refrigerated).
- Label test cuvettes or sample cups.
- Pipette the specified volume of citrate buffer into the cuvette (typically 100-200 µL).
- Add the specified volume of patient plasma (typically 50-100 µL).
- Add the specified volume of D-dimer reagent (latex-coated antibodies).
- Mix gently and incubate at 37°C for the specified time (typically 5-10 minutes).
- Measure the change in absorbance/turbidity at the specified wavelength (typically 575-660 nm).
- The instrument calculates D-dimer concentration from the calibration curve.
E. Testing Procedure (ELISA Method)
- Bring all reagents to room temperature.
- Add diluted patient samples and standards to antibody-coated wells.
- Incubate for specified time (typically 1-2 hours).
- Wash wells to remove unbound material.
- Add enzyme-conjugated antibody.
- Incubate and wash again.
- Add substrate solution and incubate.
- Add stop solution.
- Read absorbance at specified wavelength (typically 405-450 nm).
- Calculate concentration from standard curve.
F. Result Interpretation
- Normal/Negative: <0.5 µg/mL FEU (or <500 ng/mL FEU)
- Elevated/Positive: ≥0.5 µg/mL FEU
- Note: Cutoff values may vary by manufacturer and clinical application.
G. Units of Measurement
- FEU (Fibrinogen Equivalent Units): Most common; approximately 2× DDU
- DDU (D-Dimer Units): Reports actual D-dimer concentration
- Conversion: 1 µg/mL FEU ≈ 0.5 µg/mL DDU
6. Quality Control
- Run two levels of QC with each batch (normal and elevated).
- Acceptable QC ranges should be established per CLSI guidelines.
- Participate in external quality assessment/proficiency testing programs.
- Monitor delta checks for repeat patients.
- Correlate with clinical presentation and other coagulation tests.
7. Reference Ranges
- The canonical adult negative threshold for D-dimer is shown in the Reference Ranges panel below.
- Age-Adjusted Cutoff (for patients >50 years): Age (years) × 10 ng/mL FEU
- Example: 70-year-old patient: cutoff = 700 ng/mL FEU
- Note: Each laboratory should establish or verify reference intervals.
8. Clinical Significance
Elevated D-dimer may indicate:
- Deep vein thrombosis (DVT)
- Pulmonary embolism (PE)
- Disseminated intravascular coagulation (DIC)
- Recent surgery or trauma
- Pregnancy
- Malignancy
- Inflammation or infection
- Advanced age
Clinical Applications:
- Rule-out VTE: High negative predictive value (>95%) in low-to-moderate risk patients
- Monitor DIC: Serial measurements track disease progression
- Assess thrombotic risk: In various clinical settings
9. Reporting Results
- Report result with units (µg/mL FEU or ng/mL FEU).
- Include reference range on report.
- Flag elevated results (>0.5 µg/mL FEU).
- For VTE rule-out, include comment: "Negative D-dimer has high predictive value for excluding VTE in low-to-moderate risk patients."
- Critical values: Report immediately to physician if >4.0 µg/mL FEU (suggests possible DIC or massive thrombosis).